4.4
Using the Sequence Function
EC1000SA
4-33
Table4-9. Setting Items in PROGRAM Screen (2/3)
Setting Name
Item
Operation
FREQ
AC frequency
Sets the AC frequency.
Setting range:
1.0 Hz to 550.0 Hz
Setting resolution: 0.1 Hz
AC frequency
operation type
Sets the method used to change to the specified frequency.
Select one of the following three options.
•
CONST: Output specified value
•
KEEP:
Output value at end of previous step
•
SWEEP: Start by outputting value at end of previous
step, then perform a linear sweep to output the
specified value at the end of the current step
WAVE
Waveform
Selects arbitrary waveform to be output.
Setting range: SIN/SQU/ARB1 to ARB16
Waveform operation
type
Sets the method used to change to the specified waveform.
Select one of the following two options.
•
CONST: Output specified waveform
•
KEEP:
Output previous step waveform
PHS
Phase
Sets output initial phase.
Setting range:
0.0
°
to 359.9
°
Setting resolution: 0.1
°
Phase operation type
Sets the method used to change to the specified phase.
Select one of the following two options.
•
CONST: Output specified phase
•
KEEP:
Output last step phase
SYNC CODE
Step synchronized
output
This sets the code (2 bits) to be output in sync with step
execution. The sequence step transitions can checked by
monitoring the step synchronized output.
Bit 0 is output from pin 5 (DOUT5) of the external control
I/O connector and bit 1 is output from pin 6 (DOUT6).
Select one of the following four types of step synchronized
output (the H/L are displayed in the order “bit 1, bit 0”).
During Idle mode, these values are “LL”.
•
LL
•
LH
•
HL
•
HH
Step synchronized
output operation type
Specifies the output method for step synchronized output.
Select one of the following two options.
•
CONST: Output specified code
•
KEEP:
Output same code as at previous step
STOP PHS
Step end phase
enable/disable
Specifies step end phase as enabled or disabled. Select one
of the following two options.
•
DISABLE (phase depends on step time)
•
ENABLE
Step end phase
Sets step end phase.
Setting range:
0.0
°
to 359.9
°
Setting resolution: 0.1
°
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...