4.1
Use as a DC Power Source
EC1000SA
4-2
4.1 Use as a DC Power Source
When using the EC1000SA as a DC voltage source, select DC - Internal signal mode (AC+DC-INT), then
enter the settings listed in
Table4-1
.
AC components (sine wave, square wave, and arbitrary waveform) can be superimposed on the DC power
source.
A protection diode may be required to be connected, depending on the load, such as a capacitor or
inductor.
Table4-1. Panel Settings When Using EC1000SA as a DC Power Source
Setting
Setting Range
Resolution Initial Value
DC voltage
100 V range
−
220.0 to +220.0 V
0.1 V
0.0 V
200 V range
−
440.0 to +440.0 V
0.1 V
0.0 V
AC voltage
100 V range
SIN/SQU
0.0 to 155.0 Vrms 0.1 Vrms
0.0 Vrms
ARB1 to ARB16
0.0 to 440.0 Vp-p 0.1 Vp-p
0.0 Vp-p
200 V range
SIN/SQU
0.0 to 310.0 Vrms 0.1 Vrms
0.0 Vrms
ARB1 to ARB16
0.0 to 880.0 Vp-p 0.1 Vp-p
0.0 Vp-p
AC voltage frequency
1.0 to 550.0 Hz
0.1 Hz
50.0 Hz
Phase when output is on
0.0 to 359.9
°
0.1
°
0.0
°
The unit for AC voltage settings differs according to the AC voltage waveform setting.
4.1.1 Connecting output terminal to load during DC output
Use the output terminal on the rear panel when outputting DC+AC or only DC from the EC1000SA.
If the load is a capacitor or inductive load, connect a diode for protection that meets the following
specifications between the output terminal of the EC1000SA and the load.
•
Maximum reverse voltage: 600 V or higher
•
Maximum forward current: 15 A or more (100 V range), 7.5 A or more (200 V range)
a) Capacitive load such as capacitor (about 50
µ
F or more) and DC-DC converter
When a capacitive load is charged, the greater the voltage changes, the more the charge
current becomes. Overshoot may be generated in the output currents depending on the load,
even if the peak current limiter of the EC1000SA operates, thus turning off the output due to
overcurrent protection.
To avoid this, gradually lower the set voltage so that decrease of the voltage changes per
certain unit time or use the sweep function of the sequence.
A block diode is required to prevent current from flowing from the load to the output terminal
of the EC1000SA even if the output voltage is lowered. Connect the diode and the load
serially as shown in
Figure 4-1
.
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...