3.2
Display and Initial Settings at Power-on
EC1000SA
3-6
Table3-2. Settings in Memory (2/3)
Function
Setting
Initial setting
Reset
Current limiter
Peak current limiter
(positive)
100 V range
+42.0 A
√
200 V range
+21.0 A
Peak current limiter
(negative)
100 V range
−
42.0 A
200 V range
−
21.0 A
RMS current limiter
100 V range
10.5 A
200 V range
5.3 A
Setting range limit
Positive voltage
setting limit
100 V range
+220.0 V
√
200 V range
+440.0 V
Negative voltage
setting limit
100 V range
−
220.0 V
200 V range
−
440.0 V
Frequency upper limit setting limit
550.0 Hz
Frequency lower limit setting limit
1.0 Hz
External input gain
setting
External input gain
100 V range
100
200 V range
200
Select measurement
display
Measurement display
RMS
Sequence output
Step time
0.1000 s
Step operation type
Constant (CONST)
Step end phase enable/disable
Disable (DISABLE)
Step end phase
0.0
°
Step termination
End (STOP)
Jump steps
0
Jump count
1
Branch steps
0
Adjusting DC offset
AC mode
100 V range
0.0 mV
√
200 V range
0.0 mV
DC mode
100 V range
0.0 mV
200 V range
0.0 mV
System settings
Keylock
OFF
Beep
ON
LCD contrast
55
LCD display color
White(WHITE)
Time unit
s
Output on/off at power-on
OFF
External control input enable/disable
Disable(DISABLE)
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...