6.5
Status System
EC1000SA
6-54
6.5.3 Operation status
The structure of the operation status register is shown in
Figure 6-6
below.
Always 0
Under sequence execution
Not used
Under sequence hold
Not used
Not used
Not used
Not used
15
14
13
12
11
10
9
8
OPCR (operation condition register)
L
o
g
ic
a
l
O
R
OPTF (operation transition filter)
Operation status summary
Status byte (bit 7)
7
6
5
4
3
2
1
0
Not used
Not used
Not used
Not used
Not used
Not used
BUSY
Not used
7
6
5
4
3
2
1
0
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
15
14
13
12
11
10
9
8
OPER (operation event register)
OPEE (operation event enable register)
Figure 6-6. Operation Status
The transition filters convert conditions to events.
The transition filters include an NTRansition filter and a PTRansition filter. The settings and operations of
these filters are described below.
•
When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
•
When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1.
•
When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
•
When the NTR and PTR filters are both set to 0, changes in OPCR are not propagated to OPER.
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...