9.4
Signal Sources
EC1000SA
9-7
9.4 Signal Sources
The signal sources that can be selected are internal (INT), external signal input (EXT), in external
(ADD), and external synchronization (SYNC).
a) Internal signal source
(INT mode, ADD mode)
See “9.1 Output”.
b) External signal input
(EXT mode, ADD mode)
Gain setting range:
100 V range: 0.0 times to 220.0 times (initial value: 100.0)
200 V range: 0.0 times to 440.0 times (initial value: 200.0)
Gain setting resolution:
0.1 times
Gain accuracy:
±
5% (DC, or 45 Hz to 65 Hz, gain is at initial value, with rated
voltage output, no load)
Output phase to input:
In-phase
Input terminal:
BNC connector (rear panel)
(also used as the external sync signal input terminal)
Input impedance:
10 kΩ
Input voltage range:
−
2.2 V to +2.2 V
Nondestructive maximum
input voltage:
±
10 V
Frequency range:
DC to 550 Hz (sine wave)
DC to 100 Hz (other than sine wave)
c) External synchronization
(SYNC mode)
Synchronization signal source:
External sync signal or line (select one)
Synchronization frequency range:
40 Hz to 500 Hz
Input terminal:
BNC connector (rear panel)
(also used as the external signal input terminal)
Note:
External sync signal is not required when line
synchronization is selected
Input impedance:
10 kΩ
Input voltage threshold value:
TTL
Minimum pulse width:
3 ms
Nondestructive maximum
input voltage:
±
10 V
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...