6.3
Command Descriptions
EC1000SA
6-20
[SOURce:]FREQuency[:IMMediate]
This sets the output frequency. When in AC-EXT/AC-SYNC/AC+DC-EXT/AC+DC-SYNC
mode, the frequency cannot be set. ([3, “Invalid in this mode”] error occurs.)
The frequency setting ranges from 1.0 Hz to 550.0 Hz.
Syntax:
[SOURce:]FREQuency[:IMMediate]
<NR2>
Parameters:
1.0 to 550.0
Command example:
FREQ 50
Query example:
FREQ?
Response format:
<NR2>
Response example:
50.0
[SOURce:]FREQuency:LIMit:HIGH
This sets the upper limit of the frequency that can be set. However, a value that is equal to or lower
than the currently set frequency cannot be set. ([5, “Out of Limiter”] error occurs.)
The limit value setting ranges from 1.0 Hz to 550.0 Hz.
Syntax:
[SOURce:]FREQuency:LIMit:HIGH
<NR2>
Parameters:
1.0 to 550.0
Command example:
FREQ:LIM:HIGH 550
Query example:
FREQ:LIM:HIGH?
Response format:
<NR2>
Response example:
550.0
[SOURce:]FREQuency:LIMit:LOW
This sets the lower limit of the frequency that can be set. However, a value that is equal to or
higher than the currently set frequency cannot be set. ([5, “Out of Limiter”] error occurs.)
The limit value setting ranges from 1.0 Hz to 550.0 Hz.
Syntax:
[SOURce:]FREQuency:LIMit:LOW
<NR2>
Parameters:
1.0 to 550.0
Command example:
FREQ:LIM:LOW 50
Query example:
FREQ:LIM:LOW?
Response format:
<NR2>
Response example:
50.0
Содержание EC1000SA
Страница 1: ...NF Corporation PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual nbn Austria GmbH...
Страница 2: ...DA00036667 003 PROGRAMMABLE AC DC POWER SOURCE EC1000SA Instruction Manual...
Страница 31: ...2 5 Calibration EC1000SA 2 10 2 5 Calibration To calibrate the EC1000SA contact NF Corporation or our agent...
Страница 236: ...6 9 Notes on Remote Control EC1000SA 6 58...
Страница 260: ...8 5 Backup Battery EC1000SA 8 8...
Страница 280: ...9 17 External Dimensions and Weight EC1000SA 9 20...