11-6
11.5.9
Setting the SpO
2
Tone Mode
Select [
Others >>
] from the [
User Maintenance
] menu. In the popup menu, you can set [
SpO
2
Tone
] as [
Mode 1
] or
[
Mode 2
].
WARNING
The same SpO
2
tone mode shall be used for the same monitors in a single area.
11.6
Measurement Limitations
If you doubt the measured SpO
2
, check patient vital signs first. Then check the monitor and SpO
2
sensor. The following
factors may influence the accuracy of measurement:
Ambient light
Physical movement (patient and imposed motion)
Diagnostic testing
Low perfusion
Electromagnetic interference, such as MRI environment
Electrosurgical units
Dysfunctional hemoglobin, such as carboxyhemoglobin (COHb) and methemoglobin (MetHb)
Presence of certain dyes, such as methylene and indigo carmine
Inappropriate positioning of the SpO
2
sensor, or use of incorrect SpO
2
sensor
Drop of arterial blood flow to immeasurable level caused by shock, anemia, low temperature or vasoconstrictor.
11.7
Masimo Information
Masimo Patents
This posting serves as notice under 35 U.S.C.§287(a) for Masimo patents: http://www.masimo.com/patents.htm.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...