14-4
For some pressures, the parameter window may show the mean pressure only. For different pressures, their default unit
may be different. If the Art and ICP pressures are measured simultaneously, the ICP parameter area will display numeric
CPP, which is obtained by subtracting ICP from the Art mean.
14.6
Changing IBP Settings
14.6.1
Changing a Pressure for Monitoring
1. Select the pressure you want to change to enter its setup menu.
In the menu, there is a figure showing the current
IBP measurement connector.
2. Select [
Label
] and then select your desired label from the list. The already displayed labels cannot be selected.
Label
Description
Label
Description
PA
Pulmonary artery pressure
CVP
Central venous pressure
Ao
Aortic pressure
LAP
Left atrial pressure
UAP
Umbilical arterial pressure
RAP
Right atrial pressure
BAP
Brachial arterial pressure
ICP
Intracranial pressure
FAP
Femoral arterial pressure
UVP
Umbilical venous pressure
Art
Arterial blood pressure
LV
Left ventricular pressure
P1 to P4
Non-specific pressure label
NOTE
When two pressures are deteted having the same label, the monitor changes one pressure label to a
currently unused one.
14.6.2
Setting the Pressure Label Order
Select [
IBP Label Order Setup >>
] from the parameter setup menu to set the display order of the pressure labels. The
default display order is: Art, pArt, CVP, pCVP, ICP, PA, AO, UAP, FAP, BAP, LV, LAP, RAP, UVP, P1, P2, P3, P4. To restore the
default setting, you can select [
Defaults
] from the [
IBP Label Order Setup
] window.
14.6.3
Setting Alarm Properties
Select [
Alarm Setup >>
] from the parameter setup menu. You can set alarm properties for this parameter in the popup
menu.
14.6.4
Changing Averaging Time
The IBP value displayed on the monitor screen is the average of data collected within a specific time. The shorter the
averaging time is, the quicker the monitor responds to changes in the patient’s blood pressure. Conversely, the longer
the averaging time is, the slower the monitor responds to changes in the patient’s blood pressure, but the measurement
accuracy will be improved. For critically ill patients, selecting shorter averaging time will help in understanding the
patient’s state.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...