A-10
A.7.5
NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3,EN1060-4 and
SP10
Technique
Oscillometry
Mode of operation
Manual, Auto and STAT
Auto mode repetition intervals
1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h, 4
h, 8 h
STAT mode cycle time
5 min
Max measurement time
90 s
Measurement ranges
(mmHg)
Big
Medium
Small
Systolic:
25 to 285
25 to 230
25 to 135
Diastolic:
10 to 250
10 to 200
15 to125
Mean:
15 to 260
15 to 215
10 to 115
Accuracy
Max mean error:
±
5 mmHg
Max standard deviation: 8 mmHg
Resolution
1mmHg
Initial cuff inflation pressure range
(mmHg)
Big
Medium:
Small:
80 to 280
80 to 210
60 to 140
Default initial cuff inflation pressure
(mmHg)
Big
Medium:
Small:
160
140
90
Software overpressure protection
Big
Medium:
Small:
297±3 mmHg
240±3 mmHg
147±3 mmHg
PR
Measurement range
40 to 240 bpm
Resolution
1 bpm
Accuracy
±3bpm or ±3%, whichever is greater
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High
Big: (low limit+5) to 270
Medium: (low limit+5) to 200
Small: (low limit+5) to 135
5
Sys Low
40 to (high limit-5)
Mean High
Big: (low limit+5) to 230
Medium: (low limit+5) to 165
Small: (low limit+5) to 110
Mean Low
20 to (high limit-5)
Dia High
Big: (low limit+5) to 210
Medium: (low limit+5) to 150
Small: (low limit+5) to 100
Dia Low
10 to (high limit-5)
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...