D-3
Measurement
Alarm Message
L
I
Cause and Solution
ECG Noisy
L
A
The ECG signal is noisy. Check for any possible sources of
signal noise around the cable and electrode, and check the
patient for great motion.
ECG Artifact
L
A
Artifacts are detected on the ECG analysis lead and as a result
heart rate cannot be calculated and Asystole, Vfib and Vtac
cannot be analyzed. Check the connections of the electrodes
and leadwires and check for any possible source of
interference around the cable and electrode. Check the
patient’s condition and check the patient for great motion.
ECG High Freq. Noise
L
A
High frequency signals are detected on the ECG analysis lead.
Check for any possible source of interference around the cable
and electrode.
ECG Low Freq. Noise L
A
Low frequency signals are detected on the ECG analysis lead.
Check for any possible source of interference around the cable
and electrode.
ECG Amplitude Too
Small
L
C
The ECG amplitude didn’t reach the detected threshold. Check
for any possible source of interference around the cable and
electrode.
ECG Config. Err
L
C
ECG configuration is wrongly downloaded. Check the
downloaded configuration and re-download the correct
configuration.
Resp
Resp Disturbed
L
A
The respiration circuit is disturbed. Restart the monitor.
Temp
Temp Cal. Err
H
C
A calibration failed. Restart the monitor.
T1 Sensor Off
L
A
The Temp sensor has become detached from the patient or
the module. Check the sensor connections.
T2 Sensor Off
L
A
SpO
2
SpO
2
Sensor Off
L*
B
The SpO
2
sensor has become detached from the patient or the
module, or there is a fault with the SpO
2
sensor, or an
unspecified SpO
2
sensor has been used. Check the sensor
application site and the sensor type, and make sure if the
sensor is damaged. Reconnect the sensor or use a new sensor.
SpO
2
Sensor Fault
L
C
SpO
2
No Sensor
L
B
SpO
2
Unknown
Sensor
L
C
SpO
2
Sensor
Incompatible
L
C
SpO
2
Too Much Light
L
C
There is too much light on the SpO
2
sensor. Move the sensor to
a place with lower level of ambient light or cover the sensor to
minimize the ambient light.
SpO
2
Low Signal
L
C
The SpO
2
signal is too low or too weak. Check the patient’s
condition and change the sensor application site. If the error
persists, replace the sensor.
SpO
2
Weak Pulse
L
C
SpO
2
Interference
L
C
The SpO2 signal has been interfered. Check for any possible
sources of signal noise around the sensor and check the
patient for great motion.
SpO
2
Board Fault
L
C
There is a problem with the SpO
2
measurement board. Do not
use the module and contact your service personnel.
NIBP
NIBP Loose Cuff
L
A
The NIBP cuff is not properly connected, or there is a leak in
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...