A-9
A.7.4
PR
Alarm limit
Range (bpm)
Step (bpm)
PR High
(low limit +2) to 300
1
PR Low
15 to (high limit-2)
PR from Mindray SpO
2
Module
Measurement range
20 to 254 bpm
Resolution
1 bpm
Accuracy
±3 bpm
Refreshing rate
1 s
SpO
2
averaging time
7 s (when sensitivity is set to High)
9 s (when sensitivity is set to Medium)
11 s (when sensitivity is set to Low)
PR from Masimo SpO
2
Module
Measurement range
25 to 240 bpm
Resolution
1 bpm
Accuracy
±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate
1 s
SPO
2
averaging time
2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion PR accuracy
±3 bpm
PR from Nellcor SpO
2
Module
Measurement range
20 to 300 bpm
Resolution
1 bpm
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate
1 s
PR from IBP Module
Measurement range
25 to 350 bpm
Resolution
1 bpm
Accuracy
±1 bpm or ±1%, whichever is greater
Refreshing rate
1 s