11-3
11.4
Applying the Sensor
1. Select an appropriate sensor according to the module type, patient category and weight.
2. Remove colored nail polish from the application site.
3. Apply the sensor to the patient.
4. Select an appropriate adapter cable according to the connector type and plug this cable into the SpO
2
connector.
5. Connect the sensor cable to the adapter cable.
Tongue Sensor Placement
You can easily place the tongue sensor as shown below.
NOTE
Be sure that the sensor cable is positioned along the side of the animal’s face and body to avoid
entanglement with the animal.
11.5
Changing SpO
2
Settings
11.5.1
Accessing SpO
2
Menus
By selecting the SpO
2
parameter window or waveform area, you can access the [
SpO
2
Setup
] menu.
11.5.2
Adjusting the Desat Alarm
The desat alarm is a high level alarm notifying you of potentially life threatening drops in oxygen saturation. Select
[
Alarm Setup >>
] from the [
SpO
2
Setup
] menu. From the popup menu, you can set low alarm limit, alarm switch, and
alarm recording for [
Desat
]. When the SpO
2
value is below the desat alarm limit and the desat alarm switch is set on, the
message [
SpO
2
Desat
] is displayed.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...