5-2
5.3
Setting Default Configuration
The monitor will load the pre-set default configuration in the following cases.
The monitor restarts 120 seconds later after the power failure.
A patient is admitted.
A patient is discharged.
Patient data is cleared.
Patient category is changed.
To set default configuration:
1. Select [
Select Default Config.
>>] in the [
Manage Configuration
] menu.
2. In the [
Select Default Config.
] menu, select [
Load the Latest Config.
] or [
Load Specified Config.
].
When you select [
Load Specified Config.
], the configuration (Canine, Feline and Other (short for other animals);to be
restored is subject to the patient category. This configuration can be either a factory configuration or a saved user
configuration. Take canine as an example, select [
Default Canine Config.
] and toggle between [
Defaults
] or user
configuration(s).
NOTE
To see which configuration is restored when the monitor starts, enter the main screen to check the prompt
information at the lower part of the screen (displayed for about 10 seconds).
5.4
Saving Current Settings
Current settings can be saved as a user configuration. Up to 3 user configurations can be saved.
To save current settings:
1. Select [
Save Current Settings As >>
] in the [
Manage Configuration
] menu.
2. In the popup dialog box, enter the configuration name and then select [
Ok
].
5.5
Editing a Configuration
1. Select [
Edit Config. >>
] in the [
Manage Configuration
] menu.
2. The popup menu shows the existing configurations on the monitor. Selecting [
Config. on USB drive >>
] will show
the existing configurations on the USB drive. Select the desired configuration and then select the [
Edit
] button.
3. Select [
Alarm Setup >>
], [
Screen Setup >>
] or [
Parameter >>
] to enter the corresponding menu in which settings
can be changed. The changed items of alarm setup will be marked in red.
4. You can select [
Save
] or [
Save as
] to save the changed configuration. Select [
Save
] to overwrite the original
configuration. Select [
Save as
] to save the changed configuration in another name.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...