7-5
WARNING
When the alarm sound is switched off, the monitor will give no audible alarm tones even if a new alarm
occurs. Therefore the user should be very careful about whether to switch off the alarm sound or not.
Do not rely exclusively on the audible alarm system for monitoring. Adjustment of alarm volume to a low
level may result in a hazard to the patient. Always keep the patient under close surveillance.
7.4.4
Changing the Alarm Tone Pattern
To change the alarm tone pattern:
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Alarm Setup >>
] to enter the [
Alarm Setup
]
menu.
3. Select [
Alarm Sound
] and toggle between [
ISO
], [
Mode 1
] and [
Mode 2
].
User or factory default configurations exert no impact on the setup of the alarm tone pattern. The alarm tone pattern
remains unchanged after the monitor restarts.
7.4.5
Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm tone is silenced or turned off, the monitor issues a periodical
reminder tone. To set the reminder tones:
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Alarm Setup >>
] to enter the [
Alarm Setup
] menu.
3. Set the [
Reminder Tones
] to [
On
], [
Off
] or [
Re-alarm
]. When [
Re-alarm
] is selected, the current physiological
alarms and technical alarms marked with “
√
” will be re-generated after the [
Reminder Interval
].
To set the interval between reminder tones, select [
Reminder Interval
] and toggle between [
1 min
], [
2 min
] and [
3
min
].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm reminder tones, select [
Main
Menu
]
→
[
Alarm Setup >>
]
→
[
Others
] or the [
Alarm Setup
] QuickKey
→
[
Others
]. Then, select [
Reminder Vol
] and
toggle between [
High
], [
Medium
] and [
Low
].
7.5
Understanding the Alarm Setup Menu
Select the [
Alarm Setup
] QuickKey or [
Main Menu
]
→
[
Alarm Setup >>
] to enter the [
Alarm Setup
], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
Please refer to the
ECG
section for how to change ST alarm settings, how to change arrhythmia alarm settings and how
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...