A-12
A.7.8
CO
2
Measurement mode
Sidestream, microstream
Technique
Infrared absorption
Sidestream CO
2
Module
Standard
Meet standard of ISO 21647
CO
2
Measurement range
0 to 99 mmHg
Accuracy*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2 mmHg
±5% of the reading
±10% of the reading
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours
Resolution
1 mmHg
Sample flowrate
70 ml/min, 100 ml/min
Sample flowrate tolerance
15% or 15 ml/min, whichever is greater.
Warm-up time
Iso accuracy mode: ≤45s
Full accuracy mode: ≤10 min,
Rise time
Measured with a watertrap and a 2.5-meter sampling line
<400 ms @ 70 ml/min
<330 ms @ 100 ml/min
<300 ms @ 120 ml/min
<240 ms @ 150 ml/min
Gas sampling delay time
Measured with a watertrap and a 2.5-meter sampling line:
<4 s @ 100 ml/min
<4.5 s @ 70 ml/min
awRR measurement range
0 to 120 rpm
awRR measurement precision
±2 rpm
Apnea time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Note: The response time is the sum of the rise time and the delay time.
Effect of interference gases on CO
2
measurements
Gas
Concentration (%)
Quantitive effect*
N
2
O
≤60
±1 mmHg
Hal
≤4
Sev
≤5
Iso
≤5
Enf
≤5
Des
≤15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between
0-40mmHg.
Alarm limit
Range
Step
EtCO
2
High
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
Low
1 to (high limit - 2) mmHg
FiCO
2
High
1 to 99 mmHg
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...