9-4
9.8
Changing Resp Wave Settings
WARNING
When monitoring in manual detection mode, make sure to check the respiration detection level after you
have increased or decreased the size of the respiration wave.
In the [
Resp Setup
] menu, you can:
Select [
Gain
] and then select an appropriate setting. The bigger the gain is, the larger the wave amplitude is.
Select [
Sweep
] and then select an appropriate setting. The faster the wave sweeps, the wider the wave is.
9.9
Setting RR Source
To set RR source:
1. Enter the [
Resp Setup
] menu.
2. Select [
RR Source
] and then select a source or [
Auto
] from the dropdown list.
The dropdown list displays the currently available RR source. When you select [
Auto
], the system will automatically
select the RR source according to the priority. When the current RR source does not have valid measurement, the system
will automatically switch the [
RR Source
] to [
Auto
]. RR source switches back to the valid source if you press the
silence hardkey on the monitor’s front during an apnea alarm.
The priority of RR source is (from high to low): CO
2
measurement and impedance respiration measurement.The [
RR
Source
] settings of Resp and CO
2
, are linked.
The RR source options and description are shown in the table below.
Option
Description
Auto
RR source is automatically selected according to the priority.
CO
2
RR source is from CO
2
measurement.
ECG
RR source is from impedance respiration measurement.
9.10
Setting alarm properties
Select [
Alarm Setup >>
] from the [
Resp Setup
] menu. In the popup menu, you can set alarm properties for this
parameter.
9.11
Switching Resp Measurement On/Off
To switch Resp measurement on, select [
Imped. Resp Measure. ON
] from the [
Resp Setup
] menu. To switch Resp
measurement off, select [
Imped. Resp Measure. OFF
] from the [
Resp Setup
] menu and then select [
Yes
] from the
popup dialog box. A line is displayed in the waveform area and no numeric but the [
Measurement OFF
] message is
displayed in the parameter area.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...