22-2
Recommended cleaning agents are:
sodium hypochlorite bleach (diluted)
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
To clean your equipment, follow these rules:
1. Shut down the monitor and disconnect it from the power line.
2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
3. Clean the exterior surface of the equipment using a soft cloth dampened with the cleaner.
4. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
5. Dry your equipment in a ventilated, cool place.
22.3
Disinfecting
Disinfection may cause damage to the equipment and is therefore not recommended for this monitor unless otherwise
indicated in your hospital’s servicing schedule. Cleaning equipment before disinfecting is recommended.
The recommended disinfectants include: ethanol 70%, isopropanol 70%, Perform classic concentrate OXY (KHSO
4
solution).
CAUTION
Never use EtO or formaldehyde for disinfection.