7-4
7.4
Alarm Tone Configuration
7.4.1
Setting the Minimum Alarm Volume
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Alarm Setup >>
] to enter the [
Alarm Setup
]
menu.
3. Select [
Minimum Alarm Volume
] and toggle between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume, which is not affected by user
or factory default configurations. The setting of minimum alarm volume remains unchanged when the monitor shuts
down and restarts.
7.4.2
Changing the Alarm Volume
1. Select the [
Volume Setup
] QuickKey or the [
Alarm Setup
] QuickKey
→
[
Others
], or [
Main Menu
]
→
[
Alarm
Setup >>
]
→
[
Others
].
2. Select the appropriate volume from [
Alm Volume
]: X-10, in which X is the minimum volume, depending on the set
minimum alarm volume, and 10 is the maximum volume.
3. Select [
High Alarm Volume
] to set the volume of the high priority alarm as [
Alm 0
], [
Alm 1
] or
[
Alm 2
].
4. Select [
Reminder Vol
] to set the volume of the reminder tone as
[High
], [
Med
] or [
Low
].
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears on the screen.
7.4.3
Setting the Interval between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your desired alarm tone pattern. For
these two patterns, the interval between alarm tones identifies the alarm levels as follows:
Mode 1:
Interval between high level alarm tones:
continuous
Interval between medium level alarm tones:
5 s.
Interval between low level alarm tones:
20 s.
Mode 2:
Interval between high level alarm tones:
1 s.
Interval between medium level alarm tones:
5 s.
Interval between low level alarm tones:
20 s.
If you choose the ISO pattern, you can change the interval between alarm tones. To change the interval between alarm
tones:
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Alarm Setup >>
] to enter the [
Alarm Setup
]
menu.
3. Select [
High Alarm Interval (s)
], [
Med Alarm Interval (s)
] and [
Low Alarm Interval (s)
] in turn and then select the
appropriate settings.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...