A-11
*Measurement accuracy verification: the blood pressure measurements measured with this device are in compliance with
the American National Standard for Electronic or Automated Sphymomanometers (ANSI/AAMI SP10) in terms of mean error
and stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical patient
population.
A.7.6
Temp
Standards
Meet standard of EN12470-4
Technique
Thermal resistance
Measurement range
0 to 50
℃
(32 to 122
℉
)
Resolution
0.1
℃
Accuracy
±0.1
℃
(without probe)
Refreshing rate
1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit
Range
Step
T1/T2 High
(low limit +1) to 50
℃
0.1
℃
T1/T2 Low
0.1 to (high limit -1)
℃
TD High
0.1 to 50
℃
A.7.7
IBP
Standards
Meet standard of EN60601-2-34/IEC60601-2-34.
Technique
Direct invasive measurement
IBP
Measurement range
-50 to 300 mmHg
Resolution
1 mmHg
Accuracy
±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate
1 s
Pressure transducer
Excitement voltage
5 VDC, ±2%
Sensitivity
5 μV/V/mmHg
Impedance range
300 to 3000Ω
Alarm limit
Range (mmHg)
Step (mmHg)
Sys High
(low limit + 2) to 300
1
Mean High
Dia High
Sys Low
-50 to (high limit – 2)
Mean Low
Dia Low
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...