20-3
Select [
Contact Type
] and toggle between [
Normally Open
] and [
Normally Closed
].
[
Normally Open
]: select if your hospital’s nurse call relay contact is normally open.
[
Normally Closed
]: select if your hospital’s nurse call relay contact is normally closed.
Select [
Alm Lev
] and set the alarm level for nurse call-triggering alarms.
Select [
Alarm Cat.
] and then select the category to which the nurse call-triggering alarms belong.
Alarm conditions are indicated to nurses only when:
The nurse call system is enabled,
An alarm that meets your preset requirements occurs, and
The monitor is not in the alarm paused or silence status.
WARNING
To obtain the nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we supply. Otherwise the nurse
call function will not work and the monitor may be damaged.
Do not rely exclusively on the nurse call system for alarm notification. Remember that the most reliable
alarm notification combines audible and visual alarm indications with the patient’s clinical condition.
NOTE
If no setting is selected from [Alm Lev] or [Alarm Cat.], no nurse call signal will be triggered whatever alarms
occur.
20.5
Network Connection
20.5.1
Setting the Network Type
The monitor supports both wired and wireless network. To set the network type, you can select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required password
→
[
Network Setup >>
]
→
[
Monitor Network
Setup>>
]
. In the [
Monitor Network Setup>>
] menu, set the [
Network Type
] to [
WLAN
] or [
LAN
].
20.5.2
Wireless Network
The monitors can be connected to a wireless network via a built-in Wi-Fi module. To set the wireless network:
1.
Select [
Main Menu
]
→
[
Maintenance>>
]
→
[
User Maintenance>>
]
→
enter the required password
→
[
Network
Setup >>
].
2.
Set the [
Network Type
] to [
WLAN
].
3.
Select [
WLAN Setup >>
] to access the [
WLAN Setup
] menu.
4.
Configure the [
Network Name (SSID)
], [
Security Type
], [
User Name
] and
[Password
].
5.
Click [
Ok
] to confirm the setting.
To test the availability of the wireless network, follow this procedure:
1.
Select [
WLAN Test >>
] in the [
WLAN Setup
] menu.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...