14-2
1. Turn off the stopcock to the patient.
2. Vent the transducer to the atmospheric pressure by turning on the stopcock to the air.
3. In the setup menu for the pressure (e.g. Art), select [
Art Zero >>
]
→
[
Zero
]. During zero calibration, the [
Zero
]
button appears dimmed. It recovers after the zero calibration is completed. To zero all IBP channels, select the
[
Zero IBP
] hotkey, and then select [
Zero All Channels
] in the popup menu.
4. After the zero calibration is completed, close the stopcock to the air and open the stopcock to the patient.
NOTE
Your hospital policy may recommend that the ICP transducer is zeroed less frequently than other
transducers.
14.4
Setting Up the Pressure Measurement
1. Plug the pressure cable into the IBP connector.
2. Prepare the flush solution.
3. Flush the system to exhaust all air from the tubing. Ensure that the transducer and stopcocks are free of air bubbles.
WARNING
If air bubbles appear in the tubing system, flush the system with the infusion solution again. Air bubbles
may lead to inaccurate pressure readings.
4. Connect the pressure line to the patient catheter.
5. Position the transducer so that it is level with the heart, approximately at the level of the midaxillary line.
6. Select the appropriate label.
3-way stopcock
Pressure transducer
Pressure adapter cable
Monitor
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...