7-3
7.3.3
Flashing Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will flash every second,
and the corresponding alarm limit will also flash at the same frequency indicating the high or low alarm limit is violated.
7.3.4
Audible Alarm Tones
The alarm tone is distinct from the heart beat tone, keystroke tone and pulse tone in frequency. This monitor has three
choices of alarm tones and patterns: ISO, Mode 1 and Mode 2. For each pattern, the alarm tones identify the alarm levels
as follows:
ISO pattern:
High level alarms:
double beep.
Medium level alarms:
triple beep.
Low level alarms:
single beep.
Mode 1:
High level alarms:
high-pitched single beep.
Medium level alarms:
double beep.
Low level alarms:
low-pitched single beep.
Mode 2:
High level alarms:
high-pitched triple beep.
Medium level alarms:
double beep.
Low level alarms:
low-pitched single beep.
7.3.5
Alarm Status Symbols
Apart from the aforementioned alarm indicators, the monitor still uses the following symbols telling the alarm status:
indicates alarms are paused.
indicates alarm sound is silenced.
indicates the alarm sound is turned off.
indicates individual measurement alarms are turned off or the system is in alarm off status.
NOTE
When multiple alarms of different levels occur simultaneously, the monitor will select the alarm of the
highest level and give visual and audible alarm indications accordingly.
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...