D-4
Measurement
Alarm Message
L
I
Cause and Solution
NIBP Air Leak
L
A
the airway.
NIBP Pneumatic Leak
L
A
Check the NIBP cuff and pump for leakages.
NIBP Cuff Type
Wrong
L
A
The cuff type applied mismatches the patient category. Verify
the patient category and replace the cuff.
NIBP Air Pressure Err
L
A
An error occurred to the air pressure. Verify that the monitor
application site meets the environmental requirements and
check if there is any source that affects the air pressure.
NIBP Weak Signal
L
A
The patient’s pulse is weak or the cuff is loose. Check the
patient’s condition and change the cuff application site. If the
error persists, replace the cuff.
NIBP Signal
Saturated
L
A
The NIBP signal is saturated due to excess motion or other
sources.
NIBP Overrange
L
A
The measured NIBP value is not within the specified range.
NIBP Excessive
Motion
L
A
Check the patient’s condition and reduce the patient motion.
NIBP Cuff Overpress.
L
A
The NIBP airway may be occluded. Check the airway and
measure again.
NIBP Equip Err
H
A
An error occurred during NIBP measurement and therefore the
monitor cannot perform analysis correctly. Check the patient’s
condition and NIBP connections, or replace the cuff.
NIBP Timeout
L
A
NIBP Measure Failed
L
A
NIBP Illegally Reset
L
A
An illegal reset occurred during NIBP measurement. Check if
the airway is occluded.
IBP
YY Sensor Off
L*
A
Check the sensor connection and reconnect the sensor.
YY Disconnected
H
C
The liquid way is disconneted from the patient, or the
three-way valve is open to the air. Check the connection of the
liquid way, or check the valve is open to the patient. If the
problem remains, contact the Customer Services Dept. for
help.
YY Non-Pulsatile
L
A
The catheter may be occluded. Please flush the catheter.
YY represents an IBP label.
CO
2
CO
2
Sensor High
Temp
L
C
Check, stop using or replace the sensor.
CO
2
Sensor Low
Temp
L
C
Check, stop using or replace the sensor.
CO
2
Temp Overrange
L
C
The operating temperature of the CO2 module goes beyond
the specified range. After it restores within the specified range,
the module will restart automatically.
CO
2
Airway High
Press.
L
C
An error occurred in the airway pressure. Check the patient
connection and patient circuit, and then restart the monitor.
CO
2
Airway Low
Press.
L
C
CO
2
High Barometric
Press.
L
C
Check the CO2 connections, make sure that the monitor
application site meets the requirements, and check for special
Содержание iMEC8 Vet
Страница 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Страница 2: ......
Страница 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Страница 54: ...6 6 FOR YOUR NOTES ...
Страница 94: ...11 8 FOR YOUR NOTES ...
Страница 170: ...A 14 FOR YOUR NOTES ...
Страница 176: ...B 6 FOR YOUR NOTES ...
Страница 201: ......
Страница 202: ...PN 046 008143 00 3 0 ...