3-4 System Preparation
3.2.4
Power the System ON
CAUTION:
To ensure safe and effective system operation, you must perform daily
maintenance and checks. If the system begins to function improperly,
immediately stop scanning. If the system continues to function
improperly, fully shut down the system and contact the Mindray
Customer Service Department or a sales representative. If you use the
system in a persistent improperly functioning state, you may harm the
patient or damage the equipment.
Check before Powering ON
Check before the system is powered on:
No.
Check Item
1
The temperature, relative humidity and atmospheric pressure meet the requirements of
the operating conditions. See “2.4.3 Environmental Conditions” for details.
2
There is no condensation.
3
There is no distortion, damage or dirt on the system and peripheral devices.
If any dirt is found, cleaning shall be performed as illustrated in section “17 System
Maintenance.”
4
There are no loose screws on the monitor or control panel.
5
There is no cable damage (e.g., power cord). Maintain secure connections to the system
at all times.
6
The probes and probe cables are free from damage or stains.
See chapter “13 Probes and Biopsy” for details of probe cleaning and disinfection.
7
No miscellaneous odds and ends are attached or affixed to the control panel.
8
Ensure that all connections are free from damage and remain clear of foreign object
blockages.
There are no obstacles around the system and its air vent.
9
Probe cleaning and disinfection.
10
The entire scanning environment and field must be clean.
11
The locking mechanism of the casters works normally.
Before turning the system on, check if the system is plugged in. Press the power button (to the
left of the control panel) to power the system on.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...