6-8 Display & Cine Review
Retrospective cine duration: set the time that user taps [Retro Capture] when playing the
first frame of the image. It also refers to saving the cine or cycles retrospectively.
Procedures:
With the ECG disabled: tap [Cine] tab, and rotate [Time (Retro)] knob to adjust it.
With the ECG enabled: tap [Cine] tab. Tap [Loop Type (Retro)] to choose the type of the
saving time and the cardiac cycles. Rotate [Time (Retro)] or [Cycle (Retro)] to adjust it.
Freeze storage setting:
Press <Freeze> to freeze the image. The system takes the time of pressing the
retrospective/prospective key as the first frame of the image. The system proceeds saving the
cine.
TIP: after being frozen, the cine length of the prospective keeps the same with the real-time
prospective.
NOTE: the system ends up saving if the cine length goes beyond the maximum value.