Setup 12-17
Procedures:
1. Select the target module.
2. Click [Export] to open the [Export All Data] screen.
3. Select the path to save the data.
4. Select the exported file and type as PDP and click [OK].
12.10.3 Importing Setup Data
This function is used to import the existing setup data to the setup data memory of the system. The
system will reset and operate according to the setup preferences that were imported.
Procedures:
1. Click [Import] to open the Load Data screen.
2. Select the imported file.
3. Click [OK], a progress bar will appear and the setup data is imported to the specified path.
4. To restore the factory setup data, click [Load Factory] on the right side of the screen.
12.10.4 Probe Check
This function enables users to check if a transducer element is in malfunction, so as to evaluate
the transducer performance.
Click [Probe Check] to open the Probe Check screen. If a transducer element is in malfunction,
it is displayed as a red spot. If a transducer element functions well, it is displayed as a green
spot.
You can export the result image to the external device.
Notes: Probe check is provided for reference only, not for confirming a diagnosis.
Each time a transducer is connected to the main system or activated, the users are
recommended to implement probe check of transducer performance.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...