15-4 Acoustic Output
15.6 Acoustic Power Control
A qualified operator may use the system controls to limit the ultrasound output and to adjust the
quality of the images. There are three categories of system controls relating to output. They are:
Controls that have direct effect on the output
Controls that indirectly control the output
Controls that are receiver controls
Direct controls
It is possible to control, if necessary, the acoustic output with the “A.power” item on the soft
menu or the corresponding knob at the bottom of the soft menu. In this case, the maximum
value of the acoustic output never exceeds an MI of 1.9 and an I
SPTA.3
of 720 mW/cm
2
in any
mode of operation.
Indirect controls
The controls that indirectly affect output are the many imaging parameters. These are
operating modes, frequency, focal point positions, overall depth and PRF.
The operating mode determines whether the ultrasound beam is scanning or non-scanning.
Thermal bioeffect is closely connected to M mode, Doppler and Color mode. Acoustic
attenuation of tissue is directly related to probe frequency. The focal point is related to the
active aperture of the probe and beam width. For higher PRF (pulse repetition frequency),
the more output pulses occur over a period of time.
Receiver controls
The receiver controls (for example, gain, dynamic range and image post-processing, etc.) do
not affect output. These controls should be used, when possible, to improve the image quality
before using controls that directly or indirectly affect output.
15.7 Acoustic Output
15.7.1 Derated Ultrasonic Output Parameters
In order to determine the relevant Ultrasonic Output Parameters, a method is used which allows
for the comparison of ultrasound systems which operate at different frequencies and are focused
at different depths. This approach, called “derating” or “attenuating”, adjusts the acoustic output as
measured in a water tank to account for the effect of ultrasound propagation through tissue. By
convention, a specific average intensity attenuation value is used, which corresponds to a loss of
0.3 dB/cm/MHz. That is, the ultrasound intensity will be reduced by 0.3 dB/MHz for every
centimeter of travel from the probe. This can be expressed by the following equation:
)
10
/
3
.
0
(
10
z
f
water
atten
c
I
I
×
×
×
=
-
Where I
atten
is the attenuated intensity, I
water
is the intensity measured in a water tank (at distance z),
fc is the center frequency of the ultrasound wave (as measured in water) and z is the distance from
the probe. The equation for attenuating pressure values is similar except that the attenuation
coefficient is 0.15 dB/cm/MHz, or one-half the intensity coefficient. The intensity coefficient is
double the pressure coefficient because intensity is proportional to the square of pressure.
The attenuation coefficient chosen, 0.3 dB/cm/MHz, is significantly lower than any specific solid
tissue in the body. This value was chosen to account for fetal examinations. In early trimester
ultrasound fetal examinations, there may be a significant fluid path between the probe and the
fetus, and the attenuation of fluid is very small. Therefore the attenuation coefficient was lowered
to account for this.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...