1-8 Safety Precautions
4. To avoid damaging the system, do not use it in the following environments:
Locations exposed to direct sunlight
Locations subject to sudden changes in environmental temperature
Dusty locations
Locations subject to vibration
Locations near heat generators
Locations with high humidity.
5. Turn the system ON only after the power has been turned OFF for a period of time. If the
system is turned ON immediately after being turned OFF, the system may not reboot
properly and could malfunction.
6. Press the <Freeze> key to freeze an image or turn off the system power before
connecting or disconnecting a probe. Failure to do so can lead to the system and/or
probe being damaged.
7. Remove the ultrasound gel from the face of the probe when the examination is
complete. Water in the gel may enter the acoustic lens and adversely affect the
performance and safety of the probe.
8. You should back up the system properly to a secure external storage media,
including system configuration, settings and patient data. Data stored to the
system’s hard drive may be lost due to system failure, improper operation or
accident.
9. Do not apply external force to the control panel. This can lead to the system being
damaged.
10. If the system is used in a small room, the room temperature may rise. Provide
proper ventilation and free air exchange.
11. To dispose of the system or any part thereof, contact the Mindray Customer Service
Department or a sales representative. Mindray is not responsible for any system
content or accessories that have been discarded improperly. Mindray is not
responsible for any system content or accessories that have been discarded
improperly.
12. Electrical and mechanical performance may be degraded due to long usage (such
as current leakage or distortion and abrasion). Image sensitivity and precision may
also deteriorate. To ensure optimal system operation, it is recommended to maintain
the system under a Mindray service agreement.
13. The replaceable fuse is inside the chassis. Refer replacement operations to Mindray
service engineers or engineers authorized by Mindray only.
14. Do not turn OFF the system power supply during printing, file storage or invoking of
other system operations. An interrupted process may not be completed, and can
become lost or corrupted.
15. The iScape feature constructs a single extended image from a series of individual
image frames. The quality of the final image is user-dependent and requires skill to
efficiently apply the feature and technique. Exercise caution when measurements
are performed from an iScape image.
16. Ensure that the current exam date and time are the same as the system date and
time.
17. Use detachable cables as mains power breaking device.
Please read the following precautions carefully to ensure safety of the patient and the operator
when using the probes.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...