Acoustic Output 15-1
15
Acoustic Output
This section of the operator’s manual applies to the overall system, including the main unit, probes,
accessories and peripherals. This section contains important safety information for device
operators pertaining to acoustic output and how to control patient exposure through use of the
ALARA (as low as reasonably achievable) principle. This section also contains information
regarding acoustic output testing and real-time output display.
Read this information carefully before using the system.
15.1 Concerns with Bioeffects
Diagnostic ultrasound is recognized as being safe. There have been no reports of injuries to
patients caused by diagnostic ultrasound.
However, it cannot be stated categorically that ultrasound is 100% safe. Studies have revealed that
ultrasound with extremely high intensity is harmful to body tissues.
Diagnostic ultrasound technology has made a great leap forward during the last several years.
This rapid advance has generated concerns about the potential risk of bioeffects when new
applications or diagnostic technologies become available.
15.2 Prudent Use Statement
Although there are no confirmed biological effects on patients caused by exposure from present
diagnostic ultrasound instruments, the possibility exists that such biological effects may be
identified in the future. Thus ultrasound should be used in a prudent manner to provide medical
benefit to the patient. High exposure levels and long exposure times should be avoided while
acquiring the necessary clinical information.
15.3 ALARA Principle (As Low As Reasonably
Achievable)
The ALARA principle must be practiced when using ultrasound energy. Practicing ALARA ensures
that the total energy level is controlled below a low enough level at which bioeffects are not
generated while diagnostic information is being accumulated. The total energy is controlled by
output intensity and total radiation time. The output intensity necessary for examinations differs
depending on the patient and the clinical case.
Not all examinations can be performed with an extremely low level of acoustic energy. Controlling
the acoustic level at an extremely low level leads to low-quality images or insufficient Doppler
signals, adversely affecting the reliability of the diagnosis. However, increasing the acoustic power
more than necessary does not always contribute to an increase in the quality of the information
required for diagnosis, rather increasing the risk of generating bioeffects.
Users must take responsibility for the safety of patients and utilize ultrasound deliberately.
Deliberate use of ultrasound means that the output power of ultrasound must be selected based
on the ALARA principle.
Additional information regarding the concept of ALARA and the possible bioeffects of ultrasound is
available in a document from the AIUM (American Institute of Ultrasound Medicine) entitled
“Medical Ultrasound Safety.”
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...