Measurement 8-5
8.4
Measurement Accuracy
Table 1 2D Image Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±3%
Trace Area
Full Screen
Within ±7%
Circ
Full Screen
Within ±10%
Angle
Full Screen
Within ±3%
Volume
Full Screen
Within ±10 %
Distance
(iScape)
Linear probes,
convex probes,
and phased
array probes
Full Screen
Within ±5 %
Micro-convex
probe
Full Screen
Within ±10 %
Table 2 Basic Time Motion Measurements
Parameter
Range
Error
Distance
Full Screen
Within ±3 %
Time
Timeline Display
Within ±2%
Heart Rate
Timeline Display
Within ±4%
Velocity (PW
mode)
10-200cm/s (C11-3E,
L12-3E, L9-3E, DE11-3E,
V11-3HE, SC6-1E, SD8-
1E, L14-5WE, L20-5E,
SC5-1NE)
10-300cm/s (SP5-1E P7-
3E)
When angle≤60º, ≤±5%.
Velocity (CW
mode)
10-200cm/s
10-300cm/s (SP5-1E P7-
3E)
When angle≤60º, ≤±5%. (not including pencil
probe)
Table 3 3D/4D image measurements
Parameter
Range
Error
Distance
Section A/B/C
Single window display: Within ±5% (not including Smart 3D
mode)
Trace Area
Section A/B/C
Single window display: Within ±7% (not including Smart 3D
mode)
Circ
Section A/B/C
Single window display: Within ±10% (not including Smart 3D
mode)
Angle
Section A/B/C
Single window display: Within ±5% (not including Smart 3D
mode)
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...