13-30 Probes and Biopsy
3. Tighten the pinch nut of the needle-guided bracket to confirm that the needle-guided bracket is
properly installed on the transducer.
4. Adjust the dial scale to the required needle type shift.
5. Adjust the needle angle to the proper shift as required (loosen the nut first, and then tighten
the nut based on the shift you need).
6. Pull the lock pin and close the V-shaped cover to fix the lock pin in the groove of the needle
type adjusting base, so as to install the needle into the guiding hole.