Image Optimization 5-69
doctors find the sagittal view rapidly and is easy for doctors accessing the anatomical structure on
sagittal view. (such as corpus callosum, cerebellar vermis, etc).
NOTE:
Smart Planes CNS is an option, which does not support Smart 3D mode.
5.10.14.1 Basic Procedures for Smart Planes CNS
1. Acquire necessary 3D data.
2. Tap [S-Planes CNS]. There are four views detected by the system, which includes TCP, TTP, MSP
and TVP.
3. Adjust the brightness, 3D iClear and the thickness to obtain the image.
4. Tap [MSP]. There are A/B/C images appearing on the screen. The sagittal plane of the fetus
appears on window C.
Rotate A, B or C plane respectively, or rotate <4D> to adjust each plane.