5-70 Image Optimization
NOTE:
To ensure the correctness of the result, please make sure the cerebral midline on the
anatomical plane is displayed clearly and ROI enclose the whole cranial region.
5. Tap [Accept] to accept the edit to the MSP. The system recalculates the TCP, TTP and TVP
according to MSP’s position. The position and the angle for TCP, TTP and TVP appear on MSP
plane.
6. Tap [TCP]/[TTP]/[TVP] to select the plane, and rotate <M>, <PW> or <C> knob to rotate the image
plane along with X/Y/Z axis. The angle value appears on the right of the image.
7. Rotate the reference line on the MSP plane. See Chapter 5.10.14.2 Other Operations
for details.
8. Tap [Auto Measure]. Tap [Edit] to edit the measurements. See Chapter 5.10.14.2 Other Operations
for details.
9. Click [Auto Comment], the system adds the orientation and the organ comments to the desired area
according to the active ultrasound image. See Chapter 5.10.14.2 Other Operations for details.
10. Tap [Accept Result] to save the measurements to the report.
11. Add the comment and body mark on the plane. Perform the measurement, and save the single
frame/multi-frame image.
5.10.14.2 Other Operations
Parameter adjusting
Brightness
Description
Adjusting the brightness of the images.
Operation
Rotate [Brightness] to adjust the parameter.
The adjusting range is: 0-100% in increment of 2%.
Thickness
Operation
Rotate [Thickness] to adjust the parameter.
The adjusting range is: 0-30 in increment of 1.
Impacts
The bigger the value is, the more the thickness information appears.
3D iClear
Operation
Tap [3D iClear] to adjust the parameter.
The adjusting range is: 1-7 in increment of 1.
Impacts
The bigger the value of iClear is, the less the noise becomes.
Auto Comment
The system adds the orientation and the organ comments to the desired area according to the active
ultrasound image.
1. Acquire 3D data. Tap [S-Planes CNS] to enter the automatic detection of the mode.
2. Click [Auto Comment]. The comments appear on the image automatically.
Each of them is:
The orientation comments describe the location of the plane, referring to A (anterior), P
(posterior), L (Left), R (right), U (up), D (down).
Organ comments describe the position of the organ, referring to CSP (cavum septum
pellucidum), T (thalamus), CH (cerebellar hemisphere), CV (cerebellar vermis), CM (cisterna
magna), LV (lateral ventricles).
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...