System Preparation 3-5
Check the system after it is powered on
Check after the system is powered on:
No.
Check Item
1
There are no unusual sounds or smells indicating possible overheating.
2
There are no persistently displayed system error messages.
3
There is no evident excessive noise, or discontinuous, absent or black items in the B mode
image.
4
Check whether there is abnormal heat on the surface of the probe during an ultrasound
procedure. If you use a probe which is giving off excessive heat, it may burn the patient.
5
The control panel keys and knobs are fully functional.
6
The touch screen and the main monitor screens display normally depending on the system
modes and image status.
7
The exam date and time are the same as the system date and time, and are displayed
correctly.
WARNING:
If you use a probe giving off excessive heat, it may burn the patient.
If you find anything not functioning properly, this may indicate that
the system is defective. In this case, shut down the system
immediately and contact Mindray Customer Service Department or
sales representative.
3.2.5
Power the System OFF
You must follow the correct procedures to power the system off. Also, after you upgrade the
software or when the system is down, you need to power off and restart it.
To power the system off normally
Press the power button on left of the control panel to see the option:
Shut down: to power the system off normally.
Standby: to enter standby status.
Cancel: to cancel the operation.
NOTE:
1. DO NOT rush direct shutdown of the system. It may damage the data.
2. After the system is upgraded, use “Shut Down” to power the system off to make the
upgraded data effective.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...