System Preparation 3-3
3.2.3
Equipotential terminal
The symbol
represents the equipotential terminal that is used for balancing the protective
earth potentials between the system and other electrical equipment.
WARNING:
1.
Be sure to connect the equipotential wire before inserting the
power plug into the receptacle. Be sure to remove the power
plug from the receptacle before disconnecting the
equipotential wire. Failure to do so may result in electric
shock.
2.
When you connect another device to this system, you should
use the equipotential wire to connect each of equipotential
terminals. Failure to do so may result in electric shock.
3.
Connect the earth cable before turning the system ON.
Disconnect the earth cable after turning the system OFF.
Failure to do so may result in electric shock.
4.
DO NOT connect this system to outlets with the same fuses
that control the current to devices such as life-support
systems. If this system malfunctions and generates
overcurrent, or when there is an instantaneous current at
power ON, fuses of the building’s supply circuit may be
tripped.
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...