13-10 Probes and Biopsy
Contacts otherwise sterile tissue or body-space
: These transducers are considered critical and
include all intraoperative transducers. These transducers must be cleaned with an appropriate
cleaner after each use, followed by a sterilization process.
Cleaning
Please refer to the instructions in the manual and follow your hospital policy and procedures for
cleaning.
1. Wear a pair of gloves to prevent infection.
2. Disconnect the ultrasound probe from the ultrasound system. If the sheath is used, take off the
sheath and discard it.
3. Wipe away the ultrasound gel or other dirt on the surface of the probe by using a piece of
disposable lint-free soft cloth or tissue.
4. Prepare a cleaning solvent (enzymatic or neutral pH detergent, e.g., liquinox, MetriZyme) by
using distilled or softened water in accordance with the operator's manual provided by the
manufacturer.
5. Immerse the ultrasound probe fully in the cleaning solvent for at least 1 minute or a period
specified by the manufacturer.
6. Wipe and wash the probe surface gently by using a piece of lint-free soft cloth or soft sponge
until no dirt is visible. When necessary, wash the locating groove of the needle-guided bracket
and other items by using disposable cotton swabs. Avoid using a brush to wash the lens
because it may damage the probe.
7. Rinse the probe thoroughly by using a large amount of distilled or softened water (about 2
gallons) at room temperature for about 30s to remove the residual dirt and cleaning solvent.
Repeat the operation twice.
8. Dry the probe by wiping with a piece of disposable lint-free soft cloth or tissue. Do not dry the
probe by heating.
9. Inspect the probe. If visible dirt still exists, repeat the preceding steps to wash the probe until it
is all clean.
10. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If
such defects exist, the probe has reached the end of its service life. In this case, stop using it
and contact the Mindray service department.
Low-level disinfection of a non-critical probe
CAUTION:
Use protective eyewear when disinfecting using sprays.
1. Wear a pair of gloves to prevent infection.
2. Clean the probe thoroughly in accordance with the cleaning procedure before disinfection.
3. Disinfect the probe by using an appropriate low-level disinfectant. If you use a wet wipe for
disinfection, refer to the wiping method specified by the supplier and the wiping duration
specified in the operator's manual provided by the manufacturer. If you use a disinfectant spray,
spray the disinfectant on a piece of disposable lint-free soft cloth and wipe the probe. For
details about the wiping duration, see the operator's manual provided by the manufacturer.
4. Wipe away the residual disinfectant on the probe by using a piece of lint-free soft cloth soaked
with distilled water. Wipe three times. Or rinse the probe thoroughly by using a large amount of
distilled or softened water (about 2 gallons) at room temperature.
5. Dry the probe by wiping with a piece of disposable lint-free soft cloth. Do not dry the probe by
heating.
6.
Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill. If
such defects exist, the probe has reached the end of its service life. In this case, stop using it
and contact the Mindray service department.
7.
Store the probe in a cool, clean and dry environment. And repeat the cleaning and disinfection
Содержание DC-80A
Страница 2: ......
Страница 24: ......
Страница 44: ......
Страница 58: ...3 14 System Preparation Uninstalling Press the clip in the direction of the arrow to get out the holder...
Страница 59: ...System Preparation 3 15...
Страница 67: ...System Preparation 3 23...
Страница 68: ......
Страница 80: ......
Страница 299: ...Probes and Biopsy 13 19...
Страница 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Страница 324: ......
Страница 334: ......
Страница 340: ......
Страница 348: ......
Страница 352: ......
Страница 363: ...Barcode Reader B 11...
Страница 368: ......
Страница 382: ......
Страница 391: ...P N 046 014137 00 3 0...