3
INSTALLATION
44
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
3.8.3 How to install a cable probe in the tank
•
Do not wind cable probes less than 400 mm / 16" in diameter
•
Use two persons to lift the housing and the probe above the process connection.
•
Hold the instrument 1 m / 3
½
ft. above the tank.
Figure 3-35:
Wind cable probes and flexible conduits carefully
1
Do not wind cable probes less than 400 mm / 16" in diameter.
2
Do not wind the flexible conduit less than 330 mm / 13" in diameter.
1
2
WARNING!
If you bend the probe too much, you will damage the instrument and it will not measure
accurately.
Figure 3-36:
Installation of instruments with cable probes
1
>1 m / 3
½
ft.
1
HB_OPTIFLEX1300_EN_D4_080428.book Page 44 Tuesday, April 29, 2008 2:47 PM