6
OPERATION
86
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
How to open a Quick Link
•
Go to Supervisor Menu > Quick Setup
Supervisor Menu > Quick Setup
Supervisor Menu > Quick Setup
Supervisor Menu > Quick Setup.
•
Select a Quick Link with the and push buttons.
•
Press the >>>> push button.
i
This opens the menu item. You can configure the instrument.
6.4.2 Protection of the instrument settings
The Passwords
Passwords
Passwords
Passwords menu lets you change the supervisor password.
How to change the supervisor password
•
Go to Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor
Supervisor > Advanced setup > Device setup > Passwords > Supervisor.
•
Type in the new 6-character password.
i
Press the 4 buttons in any sequence.
•
Type in the new 6-character password again.
i
If the second entry is not the same as the first, the instrument will display the error
message "Password mismatch". Press > + at the same time and type in the new 6-
character password again.
•
Press > and (Esc) at the same time to exit to the "save settings" window.
•
Select Save
Save
Save
Save and press
^
.
i
The instrument will go back to operator mode.
6.4.3 Network configuration
The instrument uses HART
®
communication to send information to HART
®
-compatible
equipment. It can operate in either point-to-point or multidrop mode. The instrument will
communicate in multi-drop mode if you change the HART
®
address of output 1.
How to change from point-to-point to multidrop mode
•
Enter supervisor mode.
•
Go to Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address.
NOTE!
Make a note of the password and keep it in a safe place. If you lose the password, please contact
your supplier.
INFORMATION!
For more data, refer to Networks
on page 54.