OPERATION
6
87
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
•
Enter a value between 1 and 15 (see the caution below).
•
Press the Escape buttons(>>>> + ) until you get the save/cancel screen.
•
Select save.
•
Press
^
^
^
^
.
i
Output 1 switches to multidrop mode.
How to change from multidrop to point-to-point mode
•
Enter supervisor mode.
•
Go to Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address
Advanced Setup > Output 1 (HART) > HART Address.
•
Enter the value 0000.
•
Press the Escape buttons(>>>> + ) until you get the save/cancel screen.
•
Select save.
•
Press
^
^
^
^
.
i
Output 1 switches to point-to-point mode.
6.4.4 Linearisation
You can use the linearisation table given in function to make sure that readings are consistently
accurate.
•
Go to Supervisor > Advanced setup > Installation setup > Linearisation table
Supervisor > Advanced setup > Installation setup > Linearisation table
Supervisor > Advanced setup > Installation setup > Linearisation table
Supervisor > Advanced setup > Installation setup > Linearisation table.
•
Type in the number of reference points (up to 50 points). Press
^
.
i
This shows the linearisation table with default values.
•
Press > to type in new data. The instrument reading is given on the second line Device
Device
Device
Device
distance
distance
distance
distance.
•
Fill the tank to any given level.
•
Make an appropriate reference measurement. Type this data on the line Real distance
Real distance
Real distance
Real distance.
•
Repeat steps 3 thru 5 until all the cells in the linearisation table are completed.
•
Press
^
.
•
Press > and (Esc) at the same time to exit to the "save settings" window.
•
Select Save
Save
Save
Save and press
^
.
i
The instrument will go back to operator mode.
CAUTION!
Make sure that the HART
®
address for this instrument is different from others in the multidrop
network.
HB_OPTIFLEX1300_EN_D4_080428.book Page 87 Tuesday, April 29, 2008 2:47 PM