TECHNICAL DATA
9
143
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
Measurement limits in mm
80 is
ε
r
of water; 2.3 is
ε
r
of oil
Measurement limits in inches
80 is
ε
r
of water; 2.3 is
ε
r
of oil
Probes
Top dead
zone, A1
ε
r
= 80
Bottom dead
zone, A2
ε
r
= 80
Top dead
zone, A1
ε
r
= 2.3
Bottom dead
zone, A2
ε
r
= 2.3
[mm]
Double rod
125
10
165
50
Single rod
200
10
250
50
Coaxial
10
10
10
50
Double cable
125
10
165
50
Single cable Ø8 mm
200
10
250
50
Single cable Ø4 mm
200
10
250
50
Single cable Ø2 mm
200
10
250
50
Probes
Top dead
zone, A1
ε
r
= 80
Bottom dead
zone, A2
ε
r
= 80
Top dead
zone, A1
ε
r
= 2.3
Bottom dead
zone, A2
ε
r
= 2.3
[inches]
Double rod
4.90
0.40
6.50
1.95
Single rod
7.90
0.40
9.90
1.95
Coaxial
0.40
0.40
0.40
1.95
Double cable
4.90
0.40
6.50
1.95
Single cable Ø0.3"
7.90
0.40
9.90
1.95
Single cable Ø0.15"
7.90
0.40
9.90
1.95
Single cable 0.08"
7.90
0.40
9.90
1.95
HB_OPTIFLEX1300_EN_D4_080428.book Page 143 Tuesday, April 29, 2008 2:47 PM