6
OPERATION
80
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
C.1.18
probe end
threshold
Instruments for low dielectric
applications only. If it is difficult
to identify the probe end signal
(for example: too many parasite
signals), you can increase the
probe end threshold and the
amplification (Gain) of the
signal. For more data, refer to
Thresholds and parasite
signals
on page 94.
7 Gains, min-max: 0...+1.25 V or
-1.25
…
0 V.
Depends on
Gain.
snapshot
Starts or stops the snapshot
signal filter. Obey the
instructions in the quick setup
menu to get snapshot data. For
more data, refer to
How to use
the snapshot function to filter
parasite signals that do not
move
on page 97.
yes/no
no
C.1.19
units for
tables
Sub-menu for volume and
mass conversion operations.
C.1.19.1
table length
unit
The length unit used in the
conversion table. If "free unit"
is selected, the unit instrument
uses the unit name in menu
item C.5.1.7.
m, cm, mm, inch, ft, free unit
m
C.1.19.2
conversion
unit
The volume or mass unit used
in the conversion table. If "free
unit" is selected, the unit
instrument uses the unit name
in menu item C.5.1.9.
m3, L, US gal, GB gal, ft3, bbl,
Tons, Kg, US Tons, GB Tons,
free unit
m3
C.1.20
product
density
A value greater than 0 that is
used with a volume conversion
table to calculate mass. This
menu item is not available if
you have selected a mass unit.
0
…
20000 kg/m3
0
C.1.21
volume /
mass table
The instrument uses this table
to display volume and mass
readings. Give the number of
entries on the table. Press
^
.
Type in the height and the
corresponding volume / mass
values. For more data, refer to
How to correctly configure the
instrument to measure volume
or mass
on page 90.
Number of entries min.-max:
0
…
50
2. Table
units are
selected in
menu items
C.1.19.1
and
C.1.19.2.
C.1.22
linearisatio
n table
The instrument uses this table
to increase on-site accuracy.
Give the number of entries
plotted. Fill the tank. Make a
reference measurement and
type in the correct value next to
the instrument reading. For
more data, refer to
Linearisation
on page 87.
Number of entries min.-max:
0
…
50
2
Menu No. Function
Function description
Selection list
Default
HB_OPTIFLEX1300_EN_D4_080428.book Page 80 Tuesday, April 29, 2008 2:47 PM