3
INSTALLATION
40
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
3.7 Installation recommendations for solids
3.7.1 Nozzles on conical silos
We recommend that you prepare the installation when the silo is empty.
DANGER!
Risk of electrostatic discharge (ESD). The instrument is resistant to electrostatic discharges of
up to 15 kV (30 kV with the supplementary ESD protection option - recommended for solid
applications), but it is the fitter's and the user's responsibility to prevent ESD.
CAUTION!
Install the nozzle in the correct position to measure correctly and prevent too much bending and
traction.
Figure 3-32:
Installation recommendations for liquids
1
We recommend installation without a nozzle. If not, h
≤
50 mm / 2".
2
Radius of the tank, r
3
The end of the probe must be more than 300 mm / 12" above the tank bottom.
4
Empty space (radius, R
min
) around the probe.
5
The electromagnetic (EM) field generated by the instrument. It is also the measurement zone of the probe. Make sure
that the EM field is clear of objects and product flow.
6
Ground the tank, the product and the probe (if attached).
7
Position of the process fitting from the tank wall, r/2.
h