6.4.7 How to correctly configure the instrument to measure volume or mass
The instrument can be configured to measure volume or mass. The procedure for doing this is
given in the instrument's Quick Setup
Quick Setup
Quick Setup
Quick Setup menu.
How to create a volume or mass table.
•
Go to Supervisor > Quick setup > Setup mode > Conversion
Supervisor > Quick setup > Setup mode > Conversion
Supervisor > Quick setup > Setup mode > Conversion
Supervisor > Quick setup > Setup mode > Conversion.
NOTE!
If you move the reference point below the tank bottom, add this value when give a level for the 4
and 20 mA current output settings. If you move the reference point above the tank bottom,
subtract this value when you give a level value for the 4 and 20 mA current output settings.
CAUTION!
If the level for the 20 mA is set in the blocking distance, the instrument will not be able to use the
full current output range.
Figure 6-6:
Level measurement
1
Tank bottom offset (C.1.11)
2
Tank height (C.1.2)
3
Blocking distance (C.1.9)
4
Maximum effective measuring range
5
20 mA setting (C.3.3 or C.4.3)
6
4 mA setting (C.3.2 or C.4.2)
7
Non-measurement zone