01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
•
Increase the amplitude of the level threshold.
i
This value must be more than the incorrect signal. We recommend that you set the level
threshold at half the amplitude of the correct signal.
•
Use the same gain as that used for the instrument.
•
Save the settings.
i
The threshold increases. It ignores the parasite signal and uses the first signal it finds.
Figure 6-11:
Threshold data on the instrument display screen
1
Error message "Lost" if the instrument cannot find a signal
2
Amplitude of the threshold (in volts)
3
Gain of the threshold
4
Distance of the monitored signal from the signal converter
5
Amplitude of the monitored signal (in volts)
6
Gain of the monitored signal
Esc
1
2
4
3
5
6
INFORMATION!
If 2 or >2 products are selected in the application quicksetup menu, then the level is the first
signal that goes across the level threshold.
HB_OPTIFLEX1300_EN_D4_080428.book Page 96 Tuesday, April 29, 2008 2:47 PM