Clearance between the instrument probe and other objects in the tank
3.6.2 How to attach probes to the bottom of the tank
If the liquid is agitated or turbulent, you can attach the probe to the bottom of the tank. The
procedure to attach the probe depends on the type of probe used:
•
Weld a tube with an internal diameter of 28...30 mm / 1.1...1.2" to the bottom of the tank.
i
Make sure the tube aligns with the process connection at the top of the tank.
•
Lower the probe into the tank.
WARNING!
If the instrument has to measure the level of dangerous products (ammonia etc.), make sure
that the instrument has the METAGLAS
®
seal option.
Probe type
Empty space (radius, R
min
), around the probe
[mm]
[inches]
Coaxial
6
0
0
Double rod
7
Double cable Ø4 mm/0.15"
7
100
4
Single rod
8
Single cable Ø4 mm/0.15"
8
Single cable Ø2 mm/0.08"
8
300
12
CAUTION!
Keep the probe straight
Double rod Ø8 mm/0.3"
Figure 3-19:
How to attach a double rod probe to keep it straight