7
ATEX: SUPPLEMENTARY INSTRUCTIONS
112
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
Ambient and flange temperature
The ATEX equipment category and temperature class give the ambient temperature and related
flange temperature ranges for the instrument.
Equipment category II 1 G (EEx ia instruments only)
Equipment category II 1/2 G
Equipment category II 2 G
Temperature class
Ambient temperature
Flange temperature
[
°
C]
[
°
F]
[
°
C]
[
°
F]
T6
-20
…
+60
-4
…
+140
-20
…
+60
-4
…
+140
Temperature class
Ambient temperature
Flange temperature
[
°
C]
[
°
F]
[
°
C]
[
°
F]
T6
-40
…
+60
-40
…
+140
-20
…
+60
-4
…
+140
T5
-40
…
+75
-40
…
+167
-20
…
+60
-4
…
+140
T4
-40
…
+85
-40
…
+185
-20
…
+60
-4
…
+140
Temperature class
Ambient temperature
Flange temperature
Ø2 mm/Ø0.08" single
cable probe
All other probes
°
C
°
F
°
C
°
F
°
C
°
F
T6
-40
…
+60 -40
…
+140
-40
…
+60 -40
…
+140
-40
…
+60 -40
…
+140
-40
…
+54 -40
…
+129
-40
…
+55 -40
…
+131
-40
…
+80 -40
…
+176
T5
-40
…
+75 -40
…
+167
-40
…
+75 -40
…
+167
-40
…
+75 -40
…
+167
-40
…
+69 -40
…
+156
-40
…
+70 -40
…
+158
-40
…
+95 -40
…
+203
T4
-40
…
+85 -40
…
+185
-40
…
+85 -40
…
+185
-40
…
+85 -40
…
+185
-40
…
+77 -40
…
+171
-40
…
+80 -40
…
+176 -40
…
+110 -40
…
+230
-40
…
+70 -40
…
+158
-40
…
+75 -40
…
+167 -40
…
+135 -40
…
+275
T3
-40
…
+65 -40
…
+149
-40
…
+70 -40
…
+158 -40
…
+150 -40
…
+302
-
-
-40
…
+65 -40
…
+149 -40
…
+180 -40
…
+356
-
-
-40
…
+60 -40
…
+140 -40
…
+200 -40
…
+392
HB_OPTIFLEX1300_EN_D4_080428.book Page 112 Tuesday, April 29, 2008 2:47 PM