9.3 Measurement limits
Figure 9-3:
Measurement limits
1
A1, Top dead zone:
A1, Top dead zone:
A1, Top dead zone:
A1, Top dead zone: Distance from the flange to the top limit of measuring range. Refer to the notes and table that fol-
low.
2
A2, Bottom dead zone:
A2, Bottom dead zone:
A2, Bottom dead zone:
A2, Bottom dead zone: Length at the end of the probe, where measurement is not linear.
3
D, non measurement zone:
D, non measurement zone:
D, non measurement zone:
D, non measurement zone: Zone where measurement cannot be taken.
4
Electromagnetic (EM) field:
Electromagnetic (EM) field:
Electromagnetic (EM) field:
Electromagnetic (EM) field: The EM field generated by the instrument. Keep the EM field clear of objects and product
flow. Refer to Installation recommendations for liquids
Installation recommendations for liquids
Installation recommendations for liquids
Installation recommendations for liquids in this document for the empty space (R
min
) around the probe.
5
Gas (Air)
Gas (Air)
Gas (Air)
Gas (Air)
6
Product 1
Product 1
Product 1
Product 1
7
L, Probe length:
L, Probe length:
L, Probe length:
L, Probe length: Length specified by customer in the order. This is also the maximum measuring length for some probe
types in direct mode and all instruments that operate in TBF mode. refer to
Dimensions and weight
on page 144.
8
Tank Height
Tank Height
Tank Height
Tank Height
1
2
3
5
8
d
4
6
7
h
INFORMATION!
h
hh
h is the height of the nozzle. dddd is the diameter of the tank nozzle.
NOTE!
If h < d
h < d
h < d
h < d, then the top dead zone (A1) is equal to the top dead zone for the probe only. Refer to the
table that follows.
If h
h
h
h
≥ δ
, then the top dead zone (A1) is equal to the tank nozzle height plus the top dead zone for
the probe.
NOTE!
The dimensions of the tank nozzle have no effect on the top dead zone of the coaxial probe.