6
OPERATION
84
OPTIFLEX 1300 C
www.krohne.com
01/2008 • 4000172301 - HB OPTIFLEX 1300 R02 en
Distance
>4 mA setting for distance
Tank TBO + RO
3
Tank
RO
4
Interface level
>4 mA setting for interface
level
Tank TBO + RO
1
Tank
TBO - BD
5
Interface distance
>4 mA setting for interface
distance
Tank TBO + RO
1
Tank
RO
6
Interface volume
>4 mA setting for interface
volume
Max. value in the volume table
Max. value in
the volume
table
Interface mass
>4 mA setting for interface
mass
Max. value in the mass table
Max. value in
the mass table
Ullage volume
>4 mA setting for ullage
volume
Max. value in the volume table
Max. value in
the volume
table
Ullage mass
>4 mA setting for ullage mass
Max. value in the mass table
Max. value in
the mass table
Layer
>4 mA setting for layer
Tank TBO + RO
1
Tank height -
BD
7
Layer volume
>4 mA setting for layer volume
Max. value in the volume table
Max. value in
the volume
table
Layer mass
>4 mA setting for layer mass
Max. value in the mass table
Max. value in
the mass table
1
TBO = Tank Bottom Offset (C.1.9). RO = Reference Offset (C1.10).
2
TBO = Tank Bottom Offset (C.1.9). BD = Blocking Distance (C.1.11).
3
TBO = Tank Bottom Offset (C.1.9). RO = Reference Offset (C1.10). BD = Blocking Distance (C.1.11).
4
RO = Reference Offset (C1.10).
5
TBO = Tank Bottom Offset (C.1.9). Blocking Distance (C.1.11).
6
RO = Reference Offset (C1.10)
7
BD = Blocking Distance (C.1.11).
Output function
Minimum value
Maximum value
Default
HB_OPTIFLEX1300_EN_D4_080428.book Page 84 Tuesday, April 29, 2008 2:47 PM