ADM 6725 E10
KA
T
9 - 3
Super User Mode
3
4
5
6
etc. etc.
5 NUM
A:Sound Path
M4 Enter the required number of transits, in our
example 5.
Press ENTER to confirm.
A: 106 mm Diagon
Transd. Distance
M5 The calculated transducer distance is
displayed. With an odd number of transits,
´Diagon´ appears to show that the diagonal
mode is used.
(2) 5 NUM
A:Sound Path
M4* If the measurement was interrupted and the
Sound Path was changed, display M4 will also
indicate in brackets the number of transits last
used as a reminder that the positioning of the
transducers should be checked again.
9.4
Transducer Parameters in Super User Mode
If you are working in Super User Mode, the displays for editing the transducer
parameters will appear, even though the transducers were detected and recognised by
the instrument. In this case, the serial number of the transducers will be displayed and
not the message "Transducer Type, Standard".
Q3E7-314
Transducer Type
"
P17 Confirm your transducers
(example: Q3E7-314)
.
When the transducers are confirmed, FLUXUS subsequently overjumps the rest of the
display group 'Transducer Type'. The
MAIN MENU
will appear again.
However, if the transducer type 'Special Version' is selected, the transducer
parameters can be edited.
(See chapter 4.3.5)
9.5
Malfunctions during Super User Mode
Since the Super User Mode operates without any plausibility checks, malfunctions can
occur under certain circumstances, which may result in an automatic switching-off of
the instrument or in a crash of the internal software.
For example, if you entered 0 (zero) as the number of sound paths in display M2
'Reflection Meas.', or if you specified an outside diameter of 0.1 mm.
In any such cases:
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...