9 - 2
ADM 6725 E10
KA
T
Super User Mode
IS PASSIVE NOW
SUPERUSER MODE
After entering the a.m. HotCode, this message
appears which indicates that the Super User
Mode is now passive.
Confirm by pressing ENTER.
(The MAIN MENU will appear in standard mode.)
9.2 Super User Mode - Things worth
mentioning
•
There are no plausibility checks when parameters are being entered.
•
There are no limit checks, and there is no control over the parameters so that they
are kept within the limits of physical laws and the specification.
•
The cut-off flow velocity is not active.
•
The choice on how, during measurement, the sound path should go through the
pipe, is extended. The display M2 'Reflection Meas.' of program branch
MEASURING changes accordingly
(see the next chapter 9.3)
.
9.3
Decide which Sound Path to use
Measuring
par
>MEA<
opt
sf
M1
After going through the program branch
MEASURING in the usual way ...
!
1 NUM
Sound Path
M4
... the display M4 will appear.
Now, a number must be entered (in our example "1") which represents the number of
transit paths that the ultrasonic waves travel through the medium in the pipe. This
number characterises the sound path.
Any
odd
number requires diagonal mounting of the transducers on different sides of
the pipe.
Any
even
number requires installation of the transducers so that the reflection
method can be used.
(See chapter 5.2.6 'Transducer Mounting'.)
Synopsis
Transducer installation
in diagonal mode
Transducer installation
in reflection mode
Figure=number
of transits
Sound path
Figure=number
of transits
Sound path
1
2
Содержание FLUXUS ADM 6725
Страница 1: ...Operating Instructions RESET INIT 3x OFF ON PORTABLE ULTRASONIC FLOWMETER FLUXUS ADM 6725...
Страница 4: ...ADM 6725 E10 KAT...
Страница 14: ...1 4 ADM 6725 E10 KAT Functional Description This page has been left empty intentionally...
Страница 28: ...2 14 FLUXUS ADM 6725 E10 KAT Description of Flowmeter This page has been left empty intentionally...
Страница 69: ...ADM 6725 E10 KAT 6 9 Program Branch OUTPUT OPTIONS 2700 s Storage Rate O11 6a Next display e g current loop...
Страница 70: ...6 10 ADM 6725 E10 KAT Program Branch OUTPUT OPTIONS This page has been left empty intentionally...
Страница 104: ...10 6 ADM 6725 E10 KAT Virtual Calculation Channels This page has been left blank intentionally...
Страница 112: ...11 8 ADM 6725 E10 KAT Heat Quantity Measurement This page has been left empty intentionally...
Страница 122: ...12 10 ADM 6725 E10 KAT Wall Thickness Measurement This page has been left blank intentionally...
Страница 132: ...Process Interfaces 13 10 ADM 6725 E10 KAT This page has been left blank intentionally...
Страница 138: ...14 6 ADM 6725 E10 KAT Errors This page has been left blank intentionally...